Menu/Setup/Additional functions/Limit switches/Limit switch 1 ... 8
Function Options Info
Source of data Selection
None
Sensor inputs
Binary inputs
Controller
Fieldbus signals
Mathematical functions
MRS set 1 ... 2
Factory setting
None
Specify the input or output which is to be the
data source for the limit switch.
The sources of data on offer depend on your
device version.
You can choose from connected sensors, binary
inputs, fieldbus signals, mathematical functions,
controllers and sets for measuring range
switching.
Measuring value Selection
Depends on:
Source of data
Select the measured value, see the following
table.
Measured value depending on the Source of data
Source of data Measured value
Photometer Selection
Measured value
2nd measurement value
Raw meas. current
Raw ref. current
Lamp current
Lamp voltage
Raw meas. value
Raw 2nd meas. value
pH Glass Selection
Raw value mV
pH
Temperature
pH ISFET
ORP Selection
Temperature
ORP mV
ORP %
Oxygen (amp.) Selection
Temperature
Partial pressure
Concentration liquid
Saturation
Raw value nA
(only Oxygen (amp.))
Raw value μs
(only Oxygen (opt.))
Oxygen (opt.)
Cond i Selection
Temperature
Conductivity
Resistance
(only Cond c)
Concentration
(only Cond i and Cond c 4-pol)
Cond c
Cond c 4-pol
Disinfection Selection
Temperature
Sensor current
Concentration