EasyManua.ls Logo

Endress+Hauser Liquiline CM44P - Page 115

Endress+Hauser Liquiline CM44P
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Liquiline CM44P Operation
Endress+Hauser 115
Cleaning types
The user can choose from the following cleaning types:
Standard clean
Chemoclean
Chemoclean Plus
State of cleaning: Indicates whether the cleaning program is active or not. This is for
information purposes only.
Selecting the cleaning type
1. Menu/Setup/Additional functions/Cleaning: Select a cleaning program.
You can configure 4 different types of cleaning which you can assign individually
to the inputs.
2. Cleaning type: For each cleaning program decide which type of cleaning should be
performed.
Standard cleaning
Standard cleaning involves cleaning a sensor with compressed air, for instance, as
performed with the ion-selective sensor CAS40D (connection of cleaning unit for
→  43CAS40D)
Menu/Setup/Additional functions/Cleaning/Cleaning 1 ... 4/Standard clean
Function Options Info
Cleaning time 5 to 600 s
Factory setting
10 s
Cleaning duration
The cleaning duration and interval depend on
the process and sensor.
Determine the variables empirically or based
on experience.
Define the cleaning cycle→  117.
Chemoclean
An example is the use of the CYR10 injector unit to clean pH glass sensors. (CYR10
connection →  43)
Menu/Setup/Additional functions/Cleaning/Cleaning 1 ... 4/Chemoclean
Function Options Info
Cleaning time 0 to 900 s
Factory setting
5 s
Cleaning duration
Prerinse time 0 to 900 s
Factory setting
0 s
The cleaning duration, pre-rinse and post-rinse
times and the interval depend on the process
and sensor. Determine the variables empirically
or based on experience.
Postrinse time