Operation Liquiline CM44P
122 Endress+Hauser
The calculation of the remaining capacity depends on the following factors:
Flow rate
Volume of the exchanger
Salinity of the water at the exchanger inlet
Total volume capacity of the resin
Degree of efficiency of the exchanger
Menu/Setup/Additional functions/Mathematical functions/MF1 to 8/Mode = Cation exchanger capacity
Function Options Info
Cond IEX output Read only
Cond IEX input
Flow rate
Remaining capacity
Remaining op. time
Time until %0B
1)
Configuration
Calculation Selection
Off
On
Factory setting
Off
On/off switch for the function
Unit of volume Selection
l
gal
Factory setting
l
Exchanger volume 0.0 to 99999
Factory setting
0.0
Volume of the cation exchanger
Unit depending on the option selected in Unit of
volume
TVC Resin 0.0 to 99999 eq/l or
eq/gal
Factory setting
0.0 eq/l
TVC = total volume capacity
Unit as equivalent per Unit of volume
Resin efficiency 1.0 to 100.0 %
Factory setting
100.0 %
For information on the resin efficiency, refer to
the data provided by the manufacturer of the
resin used.
Set remaining capacity Selection
Yes
No
Factory setting
No
Before you start monitoring, specify the
remaining capacity of the exchanger resin.
This value makes allowances for the re-use of
resin that has already been used.
If a value is not entered manually, 100 % is
taken as the initial value for calculating the
current remaining capacity.
Remaining capacity
Set remaining capacity = Yes
0.0 to 100.0 %
Factory setting
0.0 %
Warning limit 1.0 to 100.0 %
Factory setting
20.0 %
Specify the remaining capacity at which the
transmitter should display a diagnostic message.