Calibration Liquiline CM44P
134 Endress+Hauser
CAL/Photometer/Measurement channel (2nd measurement channel/Reference channel)/Application
adjustment
Function Options Info
Display unit Selection
Base unit
µg/l
mg/l
g/l
ppm
ppb
%
FTU
FNU
NTUeq
(user unit)
Factory setting
Base unit
Select the unit in which your measured values
should be displayed.
To convert the basic unit to the display unit, use
datasets which you edit in the CAL menu.
User unit label
Display unit = (user unit)
Free text
Meas. value format Selection
#.#
#.##
#.###
#.####
#
Factory setting
#.###
Duplicate dataset
Copy from Selection
Dataset 1 ... 5
Select the source and target dataset. Then assign
a name for the target dataset.
Copy to
Dataset name Free text
Duplicate dataset Action Execute the action.
Table
Use table Selection
Yes
No
Factory setting
No
Specify whether the table should be used or not.
Edit table Action Editor table
Create value pairs consisting of the nominal and
actual (lab-determined) absorbance for your
medium.
Calibrate table Action Add more value pairs by conducting medium of a
known nominal absorbance through the flow cell
and using the device to determine the actual
values.
To do so, follow the instructions in the firmware.
Calibration curve Action Display the calculated calibration curve.
Manual factor
Manual factor Selection
Yes
No
Factory setting
No
Manual factor 0.0000 to 99999
Factory setting
1.0000