Liquiline CM44P Diagnostics and troubleshooting
Endress+Hauser 163
DIAG/Simulation
Function Options Info
Meas. inputs Simulation of a measured value (only for
sensors)
This menu appears once for each measuring
input.
Channel : parameter
Simulation Selection
Off
On
Factory setting
Off
If you simulate the measured value, this is
indicated on the display by a simulation icon in
front of the measured value.
Main value Depends on the sensor
Set the desired simulation value.
Sim. temperature Selection
Off
On
Factory setting
Off
If you simulate the temperature measured value,
this is indicated on the display by a simulation
icon in front of the temperature.
Temperature -50.0 to +250.0 °C
(-58.0 to 482.0 °F)
Factory setting
20.0 °C (68.0 °F)
Set the desired simulation value.