Technical data Liquiline CM44P
186 Endress+Hauser
16 Technical data
16.1 Input
Measured values Photometer
Absorption (UV, color, NIR, cell growth)
Turbidity
Memosens sensors
→ Documentation of the connected sensor
Measuring ranges Photometer
OUSAF12, OUSAF21, OUSAF22, OUSAF44, OUSAF46
0 to 2.5 AU
Max. 50 OD (depending on the optical path length)
OUSAF11
0 to 3 AU
0 to 6 OD (depending on the optical path length)
OUSTF10
0 to 200 FTU
0 to 200 ppm DE
OUSBT66
0 to 4 AU
0 to 8 OD (depending on the optical path length)
Memosens sensors
→ Documentation of the connected sensor
Types of input Digital sensor inputs for sensors with Memosens protocol
Analog current inputs (optional)
Digital inputs (optional)
Analog photometer inputs
Input signal Depending on version:
Max. 2 x analog photometers
Max. 4 x binary sensor signal
2 x 0/4 to 20 mA (optional), passive, potentially isolated from one another and from the
sensor inputs
0 to 30 V
Cable specification Cable type
Cable set CUK80 for photometer sensors
Memosens data cable CYK10 or sensor fixed cable, each with cable end sleeves or M12
round-pin connector (optional, for field housing)
Cable length
All sensors except OUSBT66
Max. 100 m (330 ft)
OUSBT66
Maximum 20 m (65 ft)