Technical data Liquiline CM44P
186 Endress+Hauser
16 Technical data
16.1 Input
Measured values Photometer
• Absorption (UV, color, NIR, cell growth)
• Turbidity
Memosens sensors
→ Documentation of the connected sensor
Measuring ranges Photometer
OUSAF12, OUSAF21, OUSAF22, OUSAF44, OUSAF46
• 0 to 2.5 AU
• Max. 50 OD (depending on the optical path length)
OUSAF11
• 0 to 3 AU
• 0 to 6 OD (depending on the optical path length)
OUSTF10
• 0 to 200 FTU
• 0 to 200 ppm DE
OUSBT66
• 0 to 4 AU
• 0 to 8 OD (depending on the optical path length)
Memosens sensors
→ Documentation of the connected sensor
Types of input • Digital sensor inputs for sensors with Memosens protocol
• Analog current inputs (optional)
• Digital inputs (optional)
• Analog photometer inputs
Input signal Depending on version:
• Max. 2 x analog photometers
• Max. 4 x binary sensor signal
• 2 x 0/4 to 20 mA (optional), passive, potentially isolated from one another and from the
sensor inputs
• 0 to 30 V
Cable specification Cable type
• Cable set CUK80 for photometer sensors
• Memosens data cable CYK10 or sensor fixed cable, each with cable end sleeves or M12
round-pin connector (optional, for field housing)
Cable length
All sensors except OUSBT66
Max. 100 m (330 ft)
OUSBT66
Maximum 20 m (65 ft)