Liquiline CM44P Commissioning
Endress+Hauser 59
9 Commissioning
9.1 Function check
L
WARNING
Incorrect connection, incorrect supply voltage
Safety risks for staff and device malfunctions!
‣
Check that all connections have been established correctly in accordance with the
wiring diagram.
‣
Ensure that the supply voltage matches the voltage indicated on the nameplate.
Saving the configuration settings as a screenshot
Via the local display , you can take screenshots at any time and save them to an SD card.
1. Insert an SD card into the SD card slot in the base module.
2. Press the navigator button for at least 3 seconds.
3. In the context menu, select the Screenshot item.
The current screen is saved as a bitmap file to the SD card in the "Screenshots"
folder.
9.2 Power up
During the device startup phase, the relays and current outputs have an undefined
status for a few seconds prior to initialization. Watch out for possible effects on any
actuators which may be connected.
9.2.1 Setting the operating language
Configuring the language
If you have not already done so, close the housing cover and screw the device closed.
1. Switch on the supply voltage.
Wait for the initialization to finish.
2. Press the soft key MENU . Set your language in the top menu item.
The device can now be operated in your chosen language.