10
Cable specification
Ex specification
Output
Output signal 1x 4 to 20 mA, potentially isolated against sensor circuit
1)
2x 4 to 20 mA, potentially isolated against sensor circuit
2)
0.8 to 1.2 mA peak to peak
3)
PROFIBUS PA
4)
FOUNDATION Fieldbus
5)
Signal on alarm 21.5 mA
digital via field bus
6)
Load Max. load with an supply voltage of 24 V: 500 Ω
Max. load with an supply voltage of 30 V: 750 Ω
Output signal range
Ex specification
current output 4/20 mA
With Memosens 100 m (330 ft) max. cable length
Intrinsically safe sensor circuit with type of protection: EEx ia IIC T6
Max. output voltage U
o
Max. output current I
o
Max. output P
o
5.04 V
80 mA
112 mW
For connection to the special measuring cable CYK10
1) current output 1
2) current output 1 and current output 2 (optional)
3) only for HART communication
4) for version with PROFIBUS PA
5) for version with FOUNDATION Fieldbus
6) with Profibus PA or FOUNDATION Fieldbus only
pH adjustable, ΔpH > 0.5
ORP adjustable, ΔU > 5 mV
Oxygen adjustable
Conductivity, conductive measured none
Conductivity, inductive measured none
Temperature adjustable, minimum 2 °C (2 °F)
Intrinsically safe supply and signal circuit with type of protection:
EEx ia IIC T6 passive
Max. input voltage U
i
Max. input current I
i
Max. input P
i
Max. internal inductivity L
i
Max. internal capacity C
i
30 V
100 mA
750 mW
221 μH (output 1)
24 μH (output 2)
1.2 nF (output 1)
negligible (output 2)