Memosens The measuring point is safer with Memosens:
No-contact, inductive signal transmission, thus optimum galvanic separation
No contact corrosion
Completely water-tight
Automatic error message if data flow interrupted
Sensor calibration possible in laboratory
Look-ahead maintenance thanks to recording of sensor data, e.g.:
Total hours of operation
Hours of operation for very high or very low pH values
Hours of operation for high temperatures above 80 °C (176 °F) and 100 °C (212 °F)
Number of steam sterilisations
Software packages You can select the software functionality to suit your needs. Consequently, you can avoid a "software overload"
while still having all the required features.
You can choose from the following from these software packages:
Basic:
Standard application for the most common measuring points
Advanced:
Highest degree of accuracy (medium compensation) and safety (Cal-Timer) in critical applications
Advanced Plus:
In addition to the Advanced package, the requirements of the FDA are taken into consideration
Software
package
Features
pH /ORP Conductivity Dissolved oxygen
Basic Analog sensors
Offset and two-point
calibration
Sample calibration
Calibration with standard
buffers
Manual buffer
Temperature compensation
Temperature adjustment
Isotherm intersection
Simulation of current output
–Self-diagnosis
–Clock
Digital sensors
features like analog sensors plus:
Sensor information
Analog sensors
Sample calibration
Temperature calibration: one-point
Temperature compensation: linear,
NaCl, Ultrapure water (NaCl, HCl)
Simulation of current output
–Self-diagnosis
Concentration measurement
–Clock
Digital sensors
Slope calibration
in air (100% rH)
in water (100% air
saturated)
in air (with input of the
absolute air pressure und
of the relative humidity)
Zero point calibration
Sample calibration
Temperature adjustment
Simulation of current output
–Self-diagnosis
–Clock
Sensor information
Advanced
Software package "Basic" and also:
Analog sensors
Medium compensation
Calibration stability settings
Calibration timer
Sensor statistics
Logbooks
Data logger
Digital sensors
features like analog sensors plus:
Operating hours counter
Sterilizations counter
Analog sensors
Logbooks
Data logger
Calibration with separate
installation factor
(inductive measurement only)
Polarization detection
(conductive measurement only)
Temperature compensation via
user table
Two-point temperature
adjustment: offset and slope
Free assignment of current
outputs: main value and
temperature
Digital sensors
Polarization voltage setting
Medium compensation
Calibration stability settings
Calibration timer
Sensor statistics
Logbooks
Data logger
Operating hours counter
Sterilizations counter
Advanced
Plus
Software package "Advanced" and also:
21CFR11
–Audit trail
21CFR11
USP alarm and pre-alarm (United
States Pharmacopeia),
EP (European Pharmacopoeia)
–Audit trail
21CFR11
–Audit trail