CM42-MAA101EBE00 Calibration and adjustment
Endress+Hauser 13
6. Confirm acceptance of the new calibration data and finish the adjustment by clicking "OK".
The transmitter automatically switches to the measuring mode and the new
temperature offset is used.
You can cancel the calibration any time. No new data are then used to adjust the sensor.
2.9 Calibrate
1. Press the soft key for "CAL".
2. Select the type of calibration.
3. Enter the necessary parameters. (The parameters that have to be configured depend on the
type of calibration selected.)
4. Start the calibration.
5. Follow the instructions in the menu.
6. Decide whether to use the calibration data captured, or to abort or repeat the calibration.
After calibration, the transmitter automatically switches back to the measuring mode
and your measuring point is now ready for operation.
• If calibration is aborted using ESC, or if the calibration is faulty, the system continues to use
the original calibration data.
• A calibration error is shown as plain text on the display.
• Any offset set is automatically deleted after accepting the calibration.