CM42-MAA101EBE00 Diagnostics, troubleshooting
Endress+Hauser 25
3.8.2 Reset and factory settings
Each of the following functions causes the device to be restarted.
Factory default or customer factory default
All user-specific changes in the configuration are lost
‣ Make a backup of your configuration on an optional CopyDAT (CY42-C1).
‣ Only select "Factory default" or "Customer factory default" if you do not need the modified
settings with user-specific changes or have saved them beforehand.
Path: DIAG/Service/Simulation
Function Display Info
Current output 1
Simulation Options
•on
•off
Factory setting
off
• on:
The simulation value is output at current output 1.
The simulation value continues to be output even if
you exit the DIAG menu. If you want to switch back to
having the system output the measured value, you
have to set the "Simulation" to "Off".
• off:
The measured value, and not the simulation value, is
output at current output 1.
Simulation value 3.60 to 21.50 mA
Factory setting
10.00 mA
Current output 2
Simulation Options
•on
•off
Factory setting
off
• on:
The simulation value is output at current output 2.
The simulation value continues to be output even if
you exit the DIAG menu. If you want to switch back to
having the system output the measured value, you
have to set the "Simulation" to "Off".
• off:
The measured value, and not the simulation value, is
output at current output 2.
Simulation value 3.60 to 21.50 mA
Factory setting
10.00 mA
Path: DIAG / Service
Function Display Info
Device restart Options
•Abort action
• Device restart
Factory setting
Abort action
Restart the device and keep all the settings