CM42-MAA101EBE00 Calibration and adjustment
Endress+Hauser 7
2.2.3 Delta zeropnt.
The device determines the difference between the zero points of the last and second-last
calibration. A shift in the zero point does not alter the sensitivity of the measurement. However,
if the offset is not corrected this can falsify the measured value.
As with the slope, you can also define and monitor limit values for the offset. If the limit values
are exceeded, this means that maintenance must be performed on the sensor.
a0013594
Fig. 2: Delta slope
Blue Last calibration
Red Calibration currently valid
bDelta slope
a0013595
Fig. 3: Delta zeropnt.
a1 Zero point of the second-last calibration
a2 Zero point of the last calibration
aDelta zeropnt.