EasyManua.ls Logo

Endress+Hauser Liquistation CSF33 - Page 102

Endress+Hauser Liquistation CSF33
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Liquistation CSF33
102 Endress+Hauser
Menu/Setup/Outputs
Function Options Info
Signal slope Selection
Low-High
High-Low
Factory setting
Low-High
Preselect the level change of the signal.
Source of data Selection
None
Limit switch 1-8
Factory setting
None
Select the limit switch via which the status
of the relay is to be output. The limit
switches are configured in the menu "Setup/
Additional functions/Limit switch".
Having selected Function Diagnostics message
Signal slope Selection
Low-High
High-Low
Factory setting
Low-High
Preselect the level change of the signal.
Operating mode Selection
as assigned
Namur M
Namur S
Namur C
Namur F
Factory setting
as assigned
as assigned
If this option is selected, the diagnostics
messages which you have individually
assigned to the binary output are output via
the binary output.
Namur M to F
If you decided to use one of the Namur
classes, all the messages that are assigned to
the individual class are output via the binary
output. You can also change the Namur class
assignment for every diagnostics message.
(Menu/Setup/General settings/
Diagnostics/Device behavior or Menu/
Setup/Inputs/../Diagnostics settings/Diag.
behavior)
Attributed diagnostic
messages
Read-only list of diagnostic
messages
All the messages assigned to the relay
output appear on the display. You do not
have the option of editing the information
here.
Having selected Function Cleaning(only for version with sensors with the Memosens protocol)
Signal slope Selection
Low-High
High-Low
Factory setting
Low-High
Preselect the level change of the signal.
Assignment Selection
None
Cleaning 1-4
Factory setting
None
Use this function to choose the cleaning
instance which should be started when the
binary output is active.
10.5.2 Current outputs
Setting the current output range
Menu/Setup/General settings: select 0..20 mA or 4..20 mA.

Table of Contents

Other manuals for Endress+Hauser Liquistation CSF33

Related product manuals