Operation Liquistation CSF33
116 Endress+Hauser
MenuSetup/Inputs/Binary input
Function Entry for sample program
MODE On
Input mode External signal
-> Control of sampling functions via external signals
Having selected input variable External signal:
Operation Start cleaning
-> A pulse triggers the cleaning.
Signal slope Low-High
Factory setting: Low-High.
2. Select Menu/Setup/Additional functions and configure the binary output:
MenuSetup/Inputs/Outputs
Function Entry for sample program
Having selected input variable Binary output
Mode Event
Factory setting: Off
Having selected input variable Event:
Slope of signal Low-High
Factory setting: Low-High.
Event End of sampling
A pulse is switched when sampling has ended.
Bottle change
A pulse is switched when a bottle is changed.
3. Select Menu/Setup/Additional functions and configure the binary output:
MenuSetup/Inputs/Additional functions/Cleaning/Cleaning 1
Function Entry for sample program
Cleaning type Chemoclean Plus
Dosing chamber rising is only enabled in this way
Cleaning steps setup Table to create a time program: You define a maximum of 30
program steps which should be performed one after the other. For
each step, enter the duration [s] and the state (0="off", 1="on") for
each relay or output. You can specify the number and name of the
outputs further down in the menu. See below for program
examples.
Failsafe step setup Table view:
In the table specify the states that the relays or outputs should
assume in the event of an error.
Outputs 1
Select the number of outputs that the actuators, such as valves or
pumps, should activate. 0....4
Output label 1 Test (out 1)
You can assign a meaningful name to each output, e.g.: "assembly",
"acid", "base" etc.