Liquistation CSF33 Diagnostics and troubleshooting
Endress+Hauser 133
No. Message Factory settings Tests or remedial action
S
1)
D
2)
F
3)
321 Liquid sensor F On On Capacitance sensor adjusted incorrectly or
defective
No medium detection possible in the dosing
glass
No sampling possible
1. Clean sensor
2. Contact the Service Department
322 Read sub-program F On On Selected subprogram cannot be read from the
program memory
Create a new subprogram
323 Write sub-program F On On Subprogram created cannot be saved
1. Hardware error
2. Contact the Service Department
324 Delete sub-program F On On Selected subprogram cannot be deleted from the
program memory
Perform a software reset
325 Read subprogram list F On On Subprogram list cannot be read from the program
memory
Perform a software reset
326 Membrane pump F On On Vacuum pump defective
Motor cable broken
Contact the Service Department
327 Air-Manager F On On Air manager for compressed air distribution
defective
Photoelectric barrier defective
Cable defective
Contact the Service Department
328 Distribution arm F On On Distribution arm zero point not found during
reference run
1. Perform a distribution arm test under
Menu/Diagnostics/Systemtest/
Distribution arm
2. Contact the Service Department
329 Pump failure F On On Pump motor is drawing excess current
Contact the Service Department
330 Membrane pump F On On Vacuum pump control defective
Contact the Service Department
331 Peristaltic pump F On On Peristaltic pump defective
Motor cable broken
Contact the Service Department
332 Peristaltic pump F On On Control of peristaltic pump defective
Contact the Service Department
333 Pressure sensor F On On Medium detection not possible, sampling not
possible
Suction line not drained before sampling
Pressure sensor defective
1. Check the suction line, if necessary using
the pump test under Menu/Diagnostics/
Systemtest/Pump purge
2. Contact the Service Department