Liquistation CSF33 Diagnostics and troubleshooting
Endress+Hauser 135
No. Message Factory settings Tests or remedial action
S
1)
D
2)
F
3)
353 Overfill check F On Off Total capacity of bottle reached
No further sampling to current bottle is triggered
If desired, make changes to the sampling
program under Select sampling program
354 Bottle check F On Off No empty bottles available for current program
No further sampling
Check the program settings under Select
sampling program
355 Start time over M On Off Start time entered is in the past
Enter a new start time
356 Overfill check F On Off The total sample volume does not fit in the
sample bottle
Change the sample volume
357 Sampling faulted M On Off Sample discarded
There are too many sampling requests pending
Make changes to the sampling program
under: Select sampling program
358 Configuration F On On Program configuration does not match the current
device configuration
Adjust the configuration
359 Emptying error F On On Error during emptying
Emptying and sampling program is canceled
1. Check the connection to the FMSY1 module
2. Check the 4R module, replace it if necessary
3. Perform a software restart under Setup/
Diagnostics/Systemtest/Restart
366 Module connection F On On No communication with the actuator module
Check the internal connecting cable to the 1IF
module
370 Internal voltage F On On Internal voltage outside the valid range
Check supply voltage
373 Electronictemp. high M On Off High electronics temperature
Check ambient temperature and energy
consumption
374 Sensor check F On Off No measurement signal from sensor
1. Check sensor connection
2. Check sensor, replace if necessary
375 No 4R module F On On No connection to the 4R module
1. Check the 4R module, replace it if necessary
2. Perform a software restart under: Menu/
Diagnostics/Device restart
401 Factory reset F On On Factory reset is performed
403 Device verification M Off Off Device verification active, please wait
405 Service IP active C Off Off Service switch is switched on
The device can be addressed at 192.168.1.212.
Switch off the service switch to change to the
saved IP settings
412 Writing backup F On Off
Wait for the write process to be finished
413 Reading backup F On Off
Wait