Diagnostics and troubleshooting Liquistation CSF33
136 Endress+Hauser
No. Message Factory settings Tests or remedial action
S
1)
D
2)
F
3)
436 SD card (80%) M On Off SD card 80% full
1. Replace SD card with empty card.
2. Clear SD card.
3. Set the logbook properties to First In First
Out ((Setup/General settings/Logbooks).
437 SD card (100%) M On Off SD card 100% full. No longer possible to write to
the card.
1. Replace SD card with empty card.
2. Clear SD card.
3. Set the logbook properties to First In First
Out ((Setup/General settings/Logbooks).
438 SD card SD card not
plugged in
M On Off SD card not plugged in
1. Check SD card.
2. Replace SD card.
3. Disable logging.
460 Output below limit S On Off Reasons
Sensor in air
Air pockets in assembly
Sensor fouled
Incorrect flow to sensor
1. Check sensor installation
2. Clean sensor
3. Adjust assignment of current outputs
461 Output above limit S On Off
502 No text catalog F On On
Contact the Service Department
503 Language change M On Off Language change failed
Contact the Service Department
529 Diag. setup active C Off Off
Wait for maintenance to be finished.
530 Logbook at 80% M On Off 1. Save the logbook to the SD card and then
delete the logbook in the device
2. Set memory to ring memory
3. Deactivate logbook
531 Logbook full M On Off
532 License error M On Off
Contact the Service Department
540 Parameter save fail M On Off Storage of configuration has failed
Repeat
541 Parameter load ok M On Off Configuration successfully loaded
542 Parameter load fail M On Off Loading of configuration has failed
Repeat
543 Parameter load abort M On Off Configuration loading aborted
544 Parameter reset ok M On Off Factory default successful
545 Parameter reset fail M On Off Setting of device configuration to factory setting
has failed
903 Minimum flow F On On The flow is too low for flow-proportional sampling
1. Check the medium flow
2. Check the flowmeter
3. Check configuration under Setup/Inputs/
Current inputS:x