Liquistation CSF33 Diagnostics and troubleshooting
Endress+Hauser 137
No. Message Factory settings Tests or remedial action
S
1)
D
2)
F
3)
906 Cat.exchanger failure F On Off Invalid values for conductivity or flow
1. Check for valid measured values in the
menu of the mathematical function.
2. Check sensors.
3. Check minimum flow.
907 Cat.exchanger warning S On Off Limit values exceeded for conductivity or flow.
Possible reasons:
Resin depleted
Blocked pipe
Check application.
908 IEX capacity low M On Off The capacity of the exchange resin will soon be
exhausted.
Schedule resin regeneration or replacement.
909 IEX capacity exhausted F On Off The capacity of the exchange resin is exhausted.
Regenerate or replace the resin.
910 Limit switch S On Off Limit switch activated
920 No sample F On On No inflow during dosing process
Suction line blocked or leaking
No inflow of sample
1. Check suction line and suction strainer
2. Check inflow of sample
921 Pump bracket open F On On The pump bracket is detected as open
Pump bracket open
Reed contact defective
1. Close the pump bracket
2. Contact the Service Department
922 Armature cycle M On Off O-ring seals on sampling assembly approaching
end of operating life
Displayed under Diagnostics/Operating time
information/Inline armature
1. Replace the seals
2. Switch off monitoring under Setup/
General settings/Sampling/Diagnostics
settings/Process sealing
3. Contact the Service Department
923 Armature cycle M On On O-ring seals on sampling assembly have reached
end of operating life
Displayed under Diagnostics/Operating time
information/Inline armature
1. Replace the seals
2. Switch off monitoring under Setup/
General settings/Sampling/Diagnostics
settings/Process sealing
3. Contact the Service Department
924 Inline armature F On On Final sampling position (in the process) of
sampling assembly not reached or detected
Final position switch is damaged
Supply cables are damaged
1. Check the final position switches
2. Check the supply cables to the final position
switches
3. Contact the Service Department