Operation Liquistation CSF33
48 Endress+Hauser
Menu/Setup/General settings/Logbooks
Function Options Info
Line plotter Menu to define the graphic display
Axes Selection
Off
On
Factory setting
On
Should the axes (x, y) be displayed (On) or not
(Off)?
Orientation Selection
Horizontal
Vertical
Factory setting
Horizontal
You can choose whether the value curves should
be displayed from left to right (Horizontal) or
from top to bottom (Vertical). If you want to
display two data logbooks simultaneously, make
sure that both logbooks have the same settings
here.
X-Description Selection
Off
On
Factory setting
On
Decide whether a description should be
displayed for the axes and whether gridlines
should be shown. In addition, you can also
decide whether pitches should be displayed.
Y-Description
Grids
Pitches
X Pitch/Grid distance 10 to 50%
Factory setting
10 %
Determine the pitch.
Y Pitch/Grid distance
Remove Action This action removes the data logbook. Any data
that have not been saved are lost.
Example: New data logbook (Setup/General settings/Logbooks/Data logbooks/New)
1. Make the settings:
Logbook name
Assign a name. Example: "01".
Source of data
Select a data source. Example: Sensor connected to channel 1 (CH1).
Measured value
Select the measured value to be recorded. Example: pH value.
Scan time
Specify the time interval between two logbook entries.
Data logbook
Activate the logbook: specify the data storage method.
2. Run the ../Finished: action.
The device shows the new logbook in the list of data logbooks.
3. Select data logbook "01".
Additional display: Log time left.
4. Only for Fill up buffer:
Choose Overflow warning: On or Off.
On: The device displays a diagnostic message in the event of memory overflow.
5. Submenu Line plotter: Specify the type of graphic representation.
10.2.5 Configuring the sampling depending on the device version
The list of functions displayed depends on the device version selected with:
Vacuum pump
1)
Peristaltic pump
2)
Distribution drive
3)
Sampling assembly:
4)