Liquistation CSF33 Operation
Endress+Hauser 49
Menu/Setup/General settings/
Function Options Info
Sampling
Number of bottles Choice of all possible bottle
combinations
The bottle configuration you ordered
is preset in the device.
Bottle volume 0 to 100000 ml
Factory setting
Depends on the bottle
configuration
If continuous operation is selected for
a sampling program, there is the
danger of overfilling the bottles. Do
not forget to empty the bottles!
Distribution parking
3)
Selection
• Back
• None
Factory setting
Back
Causes the distribution arm to go to
the center at the back or remain
parked in the current position when
the device is started or the program is
ended.
Distribution reference Selection
• Pre sampling
• Pre bottle change
• Pre program start
Factory setting
Pre sampling
The distribution arm goes through a
reference point depending on the
option selected.
Pre sampling:
This means that the distribution arm
performs a reference test before each
individual sampling.
Pre bottle change:
This means that the distribution arm
performs a reference test in every
subprogram.
Pre program start:
This means that a single reference
test is performed before the program
starts.
Power failure Selection
• Resume program
• Stop program
Factory setting
Resume program
Decide how the sampler should react
when it is energized after a power
failure.
Resume program:
• Time and flow-paced
The program calculates the omitted
samples and enters them in the
logbook as failed. When the
program is restarted, it continues
where it was interrupted.
• Flow-paced
No samples are entered in the
logbook during the power failure.
When the program is restarted, it
continues where it was interrupted.
Sample retries
1), 2), 3)
0 to 3
Factory setting
0
If sampling is started and no sample is
drawn in, sampling can be repeated
up to 3 times.
Sampling delay 0 to 99 s
Factory setting
0 s
The start of the sampling cycle can be
delayed by up to 99 s. The binary
output is switched without any delay.