Operation Liquistation CSF33
50 Endress+Hauser
Menu/Setup/General settings/
Function Options Info
Liquid detection Selection
Automatic
Semi automatic
Manual
Factory setting
Automatic
Automatic
The last intake time determined is the
new purge time.
Semi automatic
In Semi automatic purge times and
intake times can be defined
separately.
Manual
In Manual the dosing time can also
be set.
Dosing volume
1), 4) 1)
20 to 350 ml
Factory setting
200 ml
4)
10 to 1000 ml
Factory setting
200 ml
1)
Adjust the dosing tube in the dosing
chamber to change the dosing
volume. The level in the bottle is
calculated using the set dosing
volume.
4)
If the sampling assembly CSA420 is
used, 10, 30 or 50 ml are the only
values that are permitted.
Conductive sensor
1)
Selection
Low sensitivity
Medium sensitivity
High sensitivity
Factory setting
Medium sensitivity
The switching behavior can be set
with the liquid detection function. Use
the high sensitivity setting if the
sample has a low conductivity, for
example.
Dosing chamber
1)
Selection
Dose without pressure (A)
Dose with pressure (B)
Factory setting
Dose without pressure (A)
Dosing with pressure e.g. in
conditions with low suction heights
and slight counterpressure or low
volumes.
Ext. Program Pause Selection
None
Binary input 1
Binary input 2
Factory setting
None
Binary input 1-2
Available only if the input is
configured as an external signal.
If activated via an external input:
no samples are taken
sample requests are recorded in the
program logbook and logged as
"nOK"
bottle levels are changed virtually
bottle changes are recorded
virtually and performed after the
break.
Ext. Sampling Hold Selection
None
Binary input 1
Binary input 2
Factory setting
None
Binary input 1-2
Available only if the input is
configured as an external signal.
If activated via an external input:
no samples are taken
no sample requests are recorded in
the program logbook
no bottle levels are changed
no bottle changes are recorded or
performed