Operation Liquistation CSF33
52 Endress+Hauser
Menu/Setup/General settings/
Function Options Info
Sample temperature 2 to 20 °C
Factory setting
4 °C
Set the sample compartment
temperature.
Cooling control Selection
• Standard operation
• Quick cool down
Factory setting
Standard operation
The temperature regulator is switched
off for a certain time if quick cool-
down is selected.
Defrosting
The automatic defrosting system is preset at the factory. The following menu items can only be viewed
and changed in the Expert mode.
Mode Selection
• Off
• On
Factory setting
On
Configuration of the automatic
defrosting function
Interval Selection
• Hourly
• Daily
• Weekly
• Monthly
Factory setting
Hourly
Select the defrosting interval
Time 00-01:00 to 00-23:59
Factory setting
00-04:00
Duration 00:01 to 02:00
Factory setting
00:05