Operation Liquistation CSF33
56 Endress+Hauser
4. Then select Export .
If you have already assigned the file name, you will be asked whether you want to
overwrite the existing setup.
5. Use OK to confirm or cancel and assign a new file name.
Your setup is saved on the SD card in the "Device" folder.
You cannot upload the exported setup to the device again. You must use the Save
setup . This is the only way you can save a setup to an SD card and reload it later on or
upload it to other devices.
Activation code
You require activation codes for:
• Additional functions
• Firmware upgrades
If your original device has activation codes, these can be found on . The corresponding
device functions are activated at the factory. You only require the codes if servicing
the device or deactivating fieldbus protocols.
1. Enter activation code: Menu/Setup/General settings/Extended setup/Data
management/Activation code.
2. Confirm your entry.
The new hardware or software function is activated and can be configured.
Functions that are enabled by an activation code:
Function Activation code beginning with
Two current outputs (BASE2-E module only) 081...
Web server
1)
351...
HART 0B1...
PROFIBUS DP 0B3...
Modbus TCP 0B8...
Modbus RS485 0B5...
EtherNet/IP 0B9...
PROFINET 0B7...
Feedforward control 220...
Chemoclean Plus 25...
Configure the mathematical function Cation exchanger capacity 301...
1) Via Ethernet socket of the Base2 module, for versions without an Ethernet fieldbus