Operation Liquistation CSF33
86 Endress+Hauser
Settings for the Advanced program
Menu/Setup/Sampling programs
Function Options Info
Setup program
New A list of all the programs created is
displayed. For this reason, it is often helpful
to add an "S" for Standard in the program
name.
Advanced
Program name Free text Use a distinct name for your sampling
program. The program name can be up to 16
characters long.
Bottle configuration Choice of all possible bottle
combinations
The ordered bottle configuration is preset or
the configuration selected in the setup is
displayed.
Bottle volume 0 to 100000 ml
Factory setting
30000 ml
20000 ml
Set the bottle volume. The preset value
depends on the bottle configuration
configured. The bottle volume is always 30 l
for individual containers. The preset value
depends on the bottle configuration. The
bottle volume is always 20 l for individual
containers.
In the case of asymmetric distribution, e.g. 6
x 3 l + 2 x 13 l, you can set the bottle volume
on the left and right in the menu items that
follow.
Start condition Selection
Immediate
Date/time
Volume
External start
External duration
Factory setting
Immediate
Immediate
The sampling program is started
immediately.
Date/time
The sampling program is started at a specific
time that can be configured.
Volume
The sampling program is started when a
certain totalized flow is reached.
External start
The sampling program is started by a pulse
at the configured binary input.
External duration
The sampling program is active as long as
the configured input has the corresponding
level
Having selected Start condition Date/time
Start date 01.01.2000 ... 31.12.2099
Factory setting
DD.MM.YYYY
Set the start date for the sampling program.
The format depends on the option
configured under general settings.
Start time 00:00:00 ... 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is
started. The format depends on the option
configured under general settings.
Having selected Start condition Volume
Start volume input Selection
No flow input configured
Binary input S:x
Current input S:x
Factory setting
No flow input configured
Select the start volume input. The binary
input or the current input must be
configured for this function. Only the inputs
configured for flow measurement are
displayed.