EasyManua.ls Logo

Endress+Hauser Liquistation CSF33 - Page 87

Endress+Hauser Liquistation CSF33
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Liquistation CSF33 Operation
Endress+Hauser 87
Menu/Setup/Sampling programs
Function Options Info
Start flow sum 1000 to 9,999,000 m
3
Factory setting
10,000 m
3
Set the start volume.
Having selected Start condition External start
Start signal input Selection
No program start input
configured
Binary input S:x
Factory setting
No program start input
configured
Select the program start input. The binary
input must be configured for this function.
Only the inputs configured as a program
start input are displayed.
Having selected Start condition External duration
Start signal input Selection
No program start input
configured
Binary input S:x
Factory setting
No program start input
configured
Select the program duration input. The
binary input must be configured for this
function. Only the inputs configured as a
program start input are displayed.
Having selected Start condition PROFIBUS DPor Modbus
Start signal input Selection
None
%0V DO 01
%0V DO 02
%0V DO 03
%0V DO 04
%0V DO 05
%0V DO 06
%0V DO 07
%0V DO 08
Factory setting
None
Select the program start input.
Stop condition
(not for external start)
Selection
Program end
Continuous
Date/time
External duration
Factory setting
Program end
Program end
The device stops sampling automatically
once it has run through the set program.
Continuous
The device runs through the set program
continuously in an infinite loop. Do not
forget to empty the bottles.
External duration
The device stops the set program if a pulse is
sent to a binary input configured accordingly.
Having selected Stop condition Date/time
Stop date 01.01.2000 ... 31.12.2099
Factory setting
DD.MM.YYYY
Set the stop date for the sampling program.
The format depends on the option
configured under general settings.
Stop time 00:00:00 ... 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is
stopped. The format depends on the option
configured under general settings.
Having selected Stop condition External signal

Table of Contents

Other manuals for Endress+Hauser Liquistation CSF33

Related product manuals