Operation Liquistation CSF33
88 Endress+Hauser
Menu/Setup/Sampling programs
Function Options Info
Stop signal input Selection
• No program start input
configured
• Binary input S:x
•
Factory setting
No program start input
configured
Select the program stop input. The binary
input must be configured for this function.
Only the inputs configured as a program stop
input are displayed.
Having selected Stop conditionPROFIBUS DPor Modbus
Stop signal input Selection
• None
• %0V DO 01
• %0V DO 02
• %0V DO 03
• %0V DO 04
• %0V DO 05
• %0V DO 06
• %0V DO 07
• %0V DO 08
Factory setting
None
Select the program stop input.
Setup subprogram
New
Programpart Use a distinct name for your subprogram.
The program name can be up to 16
characters long.
Sampling mode Selection
• Time paced CTCV
• Flow paced VTCV
• Time/flow paced CTVV
• Single sample
• Sampling table
• External signal
•
Factory setting
Flow paced VTCV
Time paced CTCV
A constant sampling volume is taken at
steady intervals.
Flow paced VTCV
A constant sampling volume is taken at
variable intervals.
Time/flow paced CTVV
• (only for version with peristaltic pump)
• A variable sampling volume is taken at
steady intervals.
Single sample
The device takes a single sample with a
specific volume.
Sampling table
The time and the sampling volume is
assigned to a certain bottle in the sampling
table.
External signal
A sample is taken when an external signal is
received.
The settings that depend on the sampling mode (time-paced, flow-paced and time/flow-paced sampling) are
listed in the "Program type: Basic" section.
Having selected Sampling modeSingle sample
Dosing volume
(for version with vacuum
pump or sampling
assembly)
Sampling volume
(for version with
peristaltic pump)
Vacuum pump: 20 to 350 ml
Peristaltic pump: 10 to 10000
ml
Sampling assembly: 10 to
1000 ml
Factory setting
Vacuum pump: 200 ml
Peristaltic pump: 100 ml
Sampling assembly: 200 ml
Depending on the version, set the dosing
volume or the sampling volume
The volume is taken from the setup in the
version with a vacuum pump or sampling
assembly.
The dosing accuracy and the repeatability of
a sample volume < 20 ml may vary
depending on the specific application