Liquistation CSF33 Operation
Endress+Hauser 89
Menu/Setup/Sampling programs
Function Options Info
Sampling volume 10 to 10000 ml
Factory setting
100 ml
Set the sampling volume.
The dosing accuracy and the repeatability of
a sample volume < 20 ml may vary
depending on the specific application
Having selected Sampling mode Sampling table
Sampling table
Assign the time and sampling volume to a certain bottle. Add a new entry via "INSERT". Delete an entry
via "DELETE". You can make a maximum of 24 entries.
Example:
• Bottle 1
• Bottle 2
Example:
• Delta (=waiting time):
01:00:00
• Delta (=waiting time):
00:10:00
• Volume: 100 ml
• Volume: 100 ml
1st sampling one hour after program start: 100 ml in bottle 1
2nd sampling 10 minutes later: 100 ml in bottle 2
The sampling table indicates: After the defined "Delta time" (column 2) the volume of column 3 will be
dosed into the bottle of column 1.
Enable subprogram Selection
• Immediate
• Individual dates
• Repeating date
• Interval
• Event
• External start
• Deactivation
Factory setting
Immediate
Immediate
The subprogram is enabled immediately.
Individual dates
Set the start and stop dates for enabling the
subprogram.
Repeating date
Set the start condition, activity time and
repetition interval for the subprogram.
Interval
Set the start condition, activity time and
inactivity time for the subprogram.
Event
The subprogram is enabled by an event. Up
to three measuring signals are linked to form
a start signal using "and"/"or" links.
External start
The subprogram is enabled by a pulse at a
binary input configured accordingly.
Deactivation
Subprogram 2 or 2+n is started as soon as
subprogram 1 is disabled. Only possible with
multiple subprograms.
Having selected Enable subprogram Event
Start condition Selection
• No delay
• Date/Time
• Time
Factory setting
Date/Time
No delay
The subprogram is started when the
program is enabled.
Date/Time
Set the start date and start time for enabling
the subprogram.
Time
Set the start time for enabling the
subprogram.
Activation event
Number of events Selection
• 1
• 2
• 3
Factory setting
1
Specify how many measuring inputs (1‐3)
you want to link to generate an activation
signal.