Liquistation CSF33 Operation
Endress+Hauser 91
Menu/Setup/Sampling programs
Function Options Info
Auto Confirm 00:01 ... 23:59
Factory setting
01:00
Having selected Activation input External start
Activation input Selection
No program part start input
configured
Binary input S:x
Factory setting
No program part start input
configured
Select the input for start of the subprogram.
The binary input must be configured for this
function. Only the configured inputs are
displayed.
Sample at enable
(not for single sample
and sampling table and
also not for "Immediate"
and event)
Selection
Yes
No
Factory setting
Yes
Specify whether the first sample should be
taken directly when the subprogram is
enabled. For example, with intervals, a
sample is taken at the start of every
activation interval.
Sample at disable Selection
Yes
No
Factory setting
No
Specify whether a sample should be taken
when the subprogram is disabled. For
example, with intervals, a sample is taken at
the end of every activation interval.
Deactivation Selection
Bottles full
Enable invalid
Deactivation with event
Factory setting
Enable invalid
Select the disable function of the
subprogram:
Bottles full
The subprogram is disabled once all the
assigned bottles have been filled.
Enable invalid
Disable via limit value
Deactivation with event
New parameter can be defined
Bottle change mode Selection
No
Yes
Factory setting
Yes
No
The bottle is changed following a disable/
enable
Yes
When the cycle is finished, the system
continues filling the last bottle.
Synchronize samplings Selection
To subprogram start
To clock
Factory setting
To subprogram start
The intervals defined in the sampling mode
are enabled when the subprogram is started.
To clock
The intervals defined in the sampling mode
are enabled after a specific time. For
example, if 30 min is entered this means
that the interval is only activated at a time of
xx:30.
--> You configure this time in the ""Offset
synchronization menu item.