Liquistation CSF34 Commissioning
Endress+Hauser 59
6. Select the Bottle change mode after number of samples or time for average samples.
With the option "Bottle change after a time", you can enter the change time and bottle
synchronization (None, 1st bottle change time, 1st time of change + bottle number). The
description for this can be found in the "Bottle synchronization" section.
With the option "Bottle change after a time", you can choose the bottle synchronization
before the start condition (None, 1st bottle change time, 1st time of change + bottle
number). The description for this can be found in the "Bottle synchronization" section.
1. For Multiple bottles enter the number of bottles the sample should be distributed over.
2. Start condition: immediately or after date/time
3. Stop condition: after program end or continuous operation.
4. Pressing the SAVE saves the program and ends data entry.
Example:
15000 ml
100 ml
144
MODE
144
2
A0029242-EN
The program can be started.