Diagnostics and troubleshooting Liquistation CSF34
154 Endress+Hauser
No. Message Factory settings Tests or remedial action
S
1)
D
2)
F
3)
349 Read program F On On Program created cannot be saved
Hardware error has occurred
‣
Contact Endress+Hauser Service.
351 Delete program F On On Selected program cannot be deleted from the
program memory
‣
Perform a software reset.
352 Read Programlist F On On Program list cannot be read from the program
memory
‣
Reset the device: Menu/Diagnostics/Device
restart
353 Overfill check F On Off Total capacity of bottle reached
No further sampling to current bottle is triggered
‣
If desired: Make changes to the sampling
program under Select sampling program.
354 Bottle check F On Off No empty bottles available for current program
No further sampling
‣
Check program settings under Select
sampling program.
355 Start time over M On Off Start time entered is in the past
‣
Enter a new start time.
356 Overfill check F On Off The total sample volume does not fit in the
sample bottle
‣
Change the sample volume.
357 Sampling faulted M On Off • Sample discarded
• There are too many sampling requests pending
‣
Make changes to the sampling program under
Select sampling program.
358 Configuration F On On Program configuration does not match the current
device configuration
‣
Adjust the configuration.
359 Emptying error F On On • Error during emptying
• Emptying and sampling program is canceled
1. Check the connection to the FMSY1
module.
2. Check the 4R module, replace it if
necessary.
3. Reset the device: Menu/Diagnostics/
Device restart
370 Internal voltage F On On Internal voltage outside the valid range
1. Check supply voltage.
2. Check inputs and outputs for short-
circuiting.
373 Electronictemp. high M On Off High electronics temperature
‣
Check ambient temperature and energy
consumption.
374 Sensor check F On Off No measurement signal from sensor
1. Check sensor connection.
2. Check the sensor, replace it if necessary.