Operation Liquistation CSF34
68 Endress+Hauser
Menu/Setup/General settings/
Function Options Info
Dosing chamber
1)
Selection
Dose without pressure (A)
Dose with pressure (B)
Factory setting
Dose without pressure (A)
Dosing with pressure e.g. in
conditions with low suction heights
and slight counterpressure or low
volumes.
Ext. program pause Selection
None
Binary input 1
Binary input 2
Factory setting
None
Binary input 1-2
Available only if the input is
configured as an external signal.
If activated via an external input:
no samples are taken
sample requests are recorded in the
program logbook and logged as
"nOK"
bottle levels are changed virtually
bottle changes are recorded
virtually and performed after the
break.
Ext. sampling hold Selection
None
Binary input 1
Binary input 2
Factory setting
None
Binary input 1-2
Available only if the input is
configured as an external signal.
If activated via an external input:
no samples are taken
no sample requests are recorded in
the program logbook
no bottle levels are changed
no bottle changes are recorded or
performed
Liquid detection
2)
Selection
Automatic
Semi automatic
Off
Factory setting
Automatic
If "Semiautomatic" is selected, the
purge times and intake times can be
defined separately.
Off:
The definition of the purge times and
intake times is completely time-
controlled.
Automatic:
The last intake time determined is the
new purge time.
Semi automatic:
If the suction heights tend to vary
greatly.
Rinse cycles
2)
0 to 3
Factory setting
0
The suction line is rinsed with the
sample up to 3 times.
Safety interlock(optional) Selection
Off
Factory setting
Off
If the peristaltic pump is opened, the
safety interlock stops all the
functions.
Diagnostics settings
Sensor fouling
1)
Warning 0 to 10
Factory setting
7
Indicates maintenance work must be
performed on the conductivity
sensors. If there is conductive fouling
between the conductivity 1 and
conductivity 2 electrode, a diagnostic
message is displayed when this level
of fouling is reached.