Liquistation CSF34 Operation
Endress+Hauser 69
Menu/Setup/General settings/
Function Options Info
Alarm 0 to 10
Factory setting
7
If there is conductive fouling between
the conductivity 1 and conductivity 2
electrode, a diagnostic message is
displayed when this level of fouling is
reached.
Pump tube life
2)
Control Selection
Off
On
Factory setting
On
Indicates the pump hose has to be
exchanged.
Warning 10 to 50 h
Factory setting
30 h
When the tube has been in operation
for this length of time, a diagnostic
message is displayed to indicate that
the tube should be replaced in time.
Alarm 10 to 50 h
Factory setting
30 h
Totalizer 00-00:00 to 49710-06:28
Factory setting
00-00:00
Operating time of the current pump
hose in days, hours and minutes
Reset Action The tube life counter is reset to 0:00
h.
Reset Action
Sample temperature (optional)
Temp. control Selection
Off
On
Factory setting
On
Switch the temperature control of the
sample compartment on or off here.
Sample temperature 2 to 20 °C
Factory setting
4 °C
Set the sample compartment
temperature.
Cooling control Selection
Standard operation
Quick cool down
Factory setting
Standard operation
The temperature regulator is switched
off for a certain time if quick cool-
down is selected.
Defrosting
The automatic defrosting system is preset at the factory. The following menu items can only be viewed
and changed in the Expert mode.
Mode Selection
Off
On
Factory setting
On
Configuration of the automatic
defrosting function
Interval Selection
Hourly
Daily
Weekly
Monthly
Factory setting
Hourly
Select the defrosting interval