Operation Liquistation CSF34
70 Endress+Hauser
Menu/Setup/General settings/
Function Options Info
Time 00-01:00 to 00-23:59
Factory setting
00-04:00
Duration 00:01 to 02:00
Factory setting
00:05
10.2.6 Extended setup
Diagnostics settings
The list of diagnostic messages displayed depends on the path selected. There are device-
specific messages, and messages that depend on what sensor is connected.
Menu/Setup/(General settings or Inputs<Sensor channel>)/Extended setup/Diagnostics settings/Diag.
behavior
Function Options Info
List of diagnostic messages
Select the message to be changed. Only then
is it possible to make the settings for this
message.
Diag. code Read only
Diagnostic message Selection
On
Off
Factory setting
Depends on the message
Deactivate the diagnostic message or activate it
again.
Deactivating means:
No error message in the measuring mode
No failure current at the current output
Failure current Selection
On
Off
Factory setting
Depends on the message
Decide whether a failure current should be
output at the current output if the diagnostic
message display is activated.
In the event of general device errors, the
failure current is output at all the current
outputs. In the event of channel-specific
errors, the failure current is only output at
the assigned current output.
Status signal Selection
Maintenance (M)
Out of specification (S)
Function check (C)
Failure (F)
Factory setting
Depends on the message
The messages are divided into different error
categories in accordance with NAMUR NE 107.
Decide whether a status signal assignment
should be changed for the application.
Diag. output Selection
None
Alarm relay
Binary output
Relay 1 to n (depends on
the device version)
Factory setting
None
Select a relay output and/or binary output to
which the diagnostic message should be
assigned.
For sensors with the Memosens protocol: A relay
output must first be configured for Diagnostics
before it is possible to assign the message to an
output.
(Menu/Setup/Outputs: Assign the Diagnostics
function and set Operating mode to as
assigned.)
Alarm relays are available, depending on the device version.