Operation Liquistation CSF34
76 Endress+Hauser
4. Then select Export .
If you have already assigned the file name, you will be asked whether you want to
overwrite the existing setup.
5. Use OK to confirm or cancel and assign a new file name.
Your setup is saved on the SD card in the "Device" folder.
You cannot upload the exported setup to the device again. You must use the Save
setup . This is the only way you can save a setup to an SD card and reload it later on or
upload it to other devices.
Activation code
Activation codes are required for:
Additional functions
Firmware upgrades
Modifications, e.g. deactivation of fieldbus protocols
If the original device has activation codes, these can be found on . The corresponding
device functions are activated at the factory. Codes are only required if servicing the
device or deactivating fieldbus protocols.
1. Enter the activation code: Menu/Setup/General settings/Extended setup/Data
management/Activation code.
2. Confirm your entry.
The new hardware or software function is activated and can be configured.
Functions that are enabled by an activation code:
Function Activation code beginning with
Second Memosens input 062...
Fieldbus deactivation when removing module 485DP/485MB
1)
0B0...
2 current outputs (BASE2-E module only) 081...
Web server
2) 3)
351...
HART 0B1...
PROFIBUS DP 0B3...
Modbus TCP
3)
0B8...
Modbus RS485 0B5...
EtherNet/IP
3)
0B9...
PROFINET 0B7...
Feedforward control 220...
Chemoclean Plus 25...
Formula
4)
321...
Heartbeat monitoring 2D1...
Heartbeat verification 2E1...
1) If module 485DP/485MB is removed with the fieldbus protocol activated, the device outputs an error
message. Enter the activation code from the inside nameplate. The fieldbus is only deactivated once you do
this. The appropriate activation code must then be entered to activate current outputs of the base module.
Additional current outputs (CM444R and CM448R only) are activated as soon as the corresponding
module is used.
2) Via Ethernet socket of the BASE2 module, for versions without an Ethernet fieldbus
3)
4) Mathematical function