Liquistation CSF34 Operation
Endress+Hauser 83
Menu/Setup/Sampling programs
Function Options Info
Bottle volume 0 to 100000 ml
Factory setting
30000 ml
20000 ml
Set the bottle volume. The preset value
depends on the bottle configuration
configured. The bottle volume is always 30 l
for individual containers. The preset value
depends on the bottle configuration. The
bottle volume is always 20 l for individual
containers.
In the case of asymmetric distribution, e.g.
6 x 3 l + 2 x 13 l or 12 x 1 l + 6 x 2 l, you can
set the bottle volume on the left and right in
the menu items that follow.
Sampling mode Selection
Time paced CTCV
Flow paced VTCV
Time/flow paced CTVV
External signal
Factory setting
Time paced CTCV
The following functions depend on the
option selected. These versions are
illustrated individually in the following
section to provide a clearer understanding of
the options.
Time paced CTCV
A constant sampling volume is taken at
steady intervals.
In "Advanced" only:
Time monitoring
(min: 00:01:00; max: 99:59:00)
Flow paced VTCV
A constant sampling volume is taken at
variable intervals.
Time/flow paced CTVV
A sampling volume adjusted to the flow rate
is taken at steady intervals. The sampling
volume is calculated based on the current
flow rate or the average value between two
samples.
External signal
Controlled via binary input.