Liquistation CSF48 Information on sensors with the Memosens protocol
107
8 Information on sensors with the Memosens protocol
The options described below are only available on devices with a least one Memosens input.
Sensors with the Memosens protocol have integrated electronics that save calibration data and
other information. The sensor data are automatically communicated to the transmitter when
the sensor is connected and are used to calculate the measured value.
Data digital sensors save include:
Manufacturer data
Serial number
–Order code
–Date of manufacture
Calibration data
Date of calibration
–Calibration values
Number of calibrations
Serial number of the transmitter used to perform the last calibration
Operating data
Date of initial commissioning
Hours of operation under extreme operating conditions
Sensor monitoring data
The specific data that are recorded and communicated to the transmitter depend on the
sensor used. Differences can also occur within a sensor type.
This causes different menu items to be displayed or hidden depending on the sensor con-
nected.
Pay attention to specific information in this manual.
Example:
The amperometric oxygen sensor COS51D cannot be sterilized. For this reason, you will not be
able to define limit values for sterilization in the diagnostics settings for this sensor. On the
other hand, these menu items are displayed for a sterilizable amperometric sensor, such as
COS22D.