Inputs: General Liquistation CSF48
108
9Inputs: General
9.1 Configuration
An input can be configured in one of two ways:
1. Configuration where a sensor is not connected
Select the appropriate channel.
From the list of sensor types, select the sensor which you want to configure.
Configure the channel as explained in following sections.
Connect the selected sensor type later on.
2. Configuration where a sensor is connected
Configure the channel as explained in following sections.
The following applies when configuring without a sensor:
Some settings require sensor communication.
You cannot make these settings if a sensor is not connected.
It is also possible to save a setup and transfer it to another device (see "Data management" in
the "General settings" section). This function might be a better option than performing a con-
figuration when a sensor is not connected.
9.2 Frequently occurring functions
Some submenus are identical for all sensor types.
These submenus are explained below so you can find a description of these menus at one central
place. Instead of repeating the description, the sensor-specific section then only contains a refe-
rence to this chapter.
9.2.1 Damping
The damping causes a floating average curve of the measured values over the time specified.
Path: Menu/Setup/Inputs/<Sensor type>
Function Options Info
Damping <Sensor type> 0 to 600 s
Factory setting
0 s
You specify the damping of the main measured value of
the connected sensor and that of the integrated
temperature sensor.
Damping temp.