Inputs: pH/ORP Incl. combi sensors Liquistation CSF48
122
Delta slope(only pH)
The device determines the difference in slope between the last calibration and the penultimate
calibration, and issues a warning or an alarm depending on the setting configured. The
difference is an indicator for the condition of the sensor. The greater the change, the greater the
wear experienced by the pH-sensitive glass membrane as a result of chemical corrosion or
abrasion.
Operating time Total operating time of the sensor
Warning limit Factory setting
10000 h
Diagnostics code and associated message text:
199 "Operating time"
Operation > 80°C
Warning limit Factory setting
10000 h
Diagnostics code and associated message text:
193 "Operating time"
Operation > 100°C
Warning limit Factory setting
10000 h
Diagnostics code and associated message text:
194 "Operating time"
Operation < -300 mV only pH or combi sensor
Warning limit Factory setting
10000 h
Diagnostics code and associated message text:
180 "Operating time"
Operation > 300 mV only pH or combi sensor
Warning limit Factory setting
10000 h
Diagnostics code and associated message text:
179 "Operating time"
Path: Menu/Setup/Inputs/pH or pH/Redox/Extended setup/Diagnostics settings
Function Options Info
Delta slope 0.10 to 10.00 mV/pH Specify your limit values for monitoring the slope
differential.
Function Options
•On
•Off
Factory setting
Off
Warning limit Factory setting
5.00 mV/pH
Diagnostics code and associated message text:
518 "Sensor calib."
Path: Menu/Setup/Inputs/pH or Redox or pH/Redox/Extended setup/Diagnostics settings
Function Options Info