EasyManua.ls Logo

Endress+Hauser Liquistation CSF48 - Page 127

Endress+Hauser Liquistation CSF48
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Liquistation CSF48 Inputs: Conductivity
127
11.1.7 Unit and format
11.1.8 Temperature compensation
Temperature coefficient = change in the conductivity per degree of temperature change:
(T) = (T
0
)(1 + (T - T
0
))
(T) ... conductivity at process temperature T
(T
0
) ... conductivity at reference temperature T
0
The temperature coefficient depends both on the chemical composition of the solution and the
temperature itself.
Path: Menu/Setup/Inputs/Conductivity
Function Options Info
Main value format Options
•Auto
•#
•#.#
•#.##
#.###
Factory setting
Auto
Specify the number of decimal places.
Cond. unit
(Operating
mode=Conductivity)
Unit
(Operating mode=Resistance)
Options
Conductivity/resistance
•Auto / Auto
μS/cm / Mm
•mS/cm / Mcm
S/cm / kcm
μS/m / km
•mS/m / m
S/m / cm
Factory setting
Auto / Auto
The picklist depends on the operating mode.
You can either choose from units for conductivity or
units for resistivity.
Since there are no options for concentration
measurement, this function is not displayed for such
measurements.
Path: Menu/Setup/Inputs/Conductivity
Function Options Info
Temp. source Options
•Sensor
•Manual
Factory setting
Sensor
Decide how you want to compensate the medium tem-
perature:
Automatically using the temperature sensor of your
sensor
Manually by entering the medium temperature
Medium temperature
(Temp. source=Manual)
-50.0 to 250.0 ˚C
(-58.0 to 482.0 ˚F)
Factory setting
25.0 ˚C (77 ˚F)
Enter the temperature of your medium.

Table of Contents

Other manuals for Endress+Hauser Liquistation CSF48

Related product manuals