Liquistation CSF48 Inputs: Conductivity
133
11.2.4 Pharmaceutical water
Here you can make settings for monitoring pharmaceutical water in accordance with the United
States Pharmacopeia (USP) or European Pharmacopeia (EP).
The uncompensated conductivity value and the temperature are measured for the limit func-
tions. The measured values are compared with the tables defined in the standards. If the limit is
exceeded, an alarm is triggered. Furthermore, you can also set a preliminary alarm (warning
limit) which signals undesired operating states before they occur.
11.2.5 Tag control
--> "Inputs: General/Frequently occurring functions" section ä 108
11.2.6 Sensor replacement
--> "Inputs: General/Frequently occurring functions" section ä 108
11.2.7 Data processing factory setting
--> "Inputs: General/Frequently occurring functions" section ä 108
11.2.8 Sensor factory setting (only CLS50D)
--> "Inputs: General/Frequently occurring functions" section ä 108
Path: Menu/Setup/Inputs/Conductivity/Extended setup/Diagnostics settings/Pharmacy-water
Function Options Info
Function Options
•Off
•EP
•USP
Factory setting
Off
The alarm values are stored in the device in accordance
with USP or EP specifications.
You define the warning limit as a % of the alarm value.
Warning limit 10.0 to 99.9 %
Factory setting
80.0 %
Diagnostics code and associated message text:
915 "USP / EP warning"
If the value exceeds the USP or EP alarm values saved in
the software, diagnostics message 914 "USP/ EP alarm"
is displayed.