Inputs: Oxygen Liquistation CSF48
138
Medium compensation (during calibration)
Calibration timer and calibration expiration date
--> "Inputs: General/Frequently occurring functions" section ä 108
12.2.6 Diagnostics settings
This menu branch is used for specifying warning limits, and for defining whether and how dia-
gnostics tools should be used.
The associated diagnostics code is displayed for every setting.
Slope
The (relative) slope characterizes the sensor condition. Decreasing values indicate electrolyte
exhaustion. You can control when the electrolyte should be replaced by specifying limit values
and the diagnostics messages these limit values trigger.
Path: Menu/Setup/Inputs/DO/Extended setup/Calib. settings
Function Options Info
Medium pressure Options
Process pressure
Air pressure
Altitude
Factory setting
Air pressure
Altitude
Medium pressure="Altitude"
-300 to 4000 m
Factory setting
0 m
Enter the altitude or the average air pressure (mutually
dependent values).
If you specify the altitude, the average air pressure is
calculated from the barometric altitude formula and vice
versa.
If you are compensating using the process pressure,
enter the pressure in your process here. The pressure is
then independent of the altitude.
Air pressureorProcess
pressure
Medium pressure="Air pres-
sure"
500 to 1200 hPa
Medium pressure="Process
pressure"
500 to 9999 hPa
Factory setting
1013 hPa
Rel. hum. (air variable) 0 to 100 %
Factory setting
100 %
Path: Menu/Setup/Inputs/DO/Extended setup/Diagnostics settings
Function Options Info
Slope 0.0 to 200.0 % Specify the limit values for slope monitoring in your
sensor.