Liquistation CSF48 Inputs: Oxygen
141
Cap sterilizations (only sterilizable, amperometric sensors)
The sterilization counters in the sensor make a distinction between the sensor and the
membrane cap currently used. If this cap is replaced, only the (cap) counter is reset.
Sterilizations (only sterilizable sensors)
--> "Inputs: General/Frequently occurring functions" section ä 108
Process check system
--> "Inputs: General/Frequently occurring functions" section ä 108
Limits operating hours
The total operating time of the sensor and its use under extreme conditions is monitored. If the
operating time exceeds the defined threshold values, the device issues a corresponding
diagnostics message.
Each sensor has a limited life expectancy which heavily depends on the operating
conditions. If you specify warning limits for operating times under extreme conditions, you
can guarantee the operation of your measuring point without any downtime by
performing maintenance tasks in time.
Path: Menu/Setup/Inputs/DO/Extended setup/Diagnostics settings
Function Options Info
Number of cap sterilizations Specify how many sterilizations may be performed with
a membrane cap before the cap has to be replaced.
The number depends heavily on the process and must be
determined individually.
Function Options
•On
•Off
Factory setting
Off
Warning limit 0 to 100
Factory setting
25
Diagnostics code and associated message text:
109 "Sterilizat. cap"