Liquistation CSF48 Inputs: Chlorine
149
Delta zero point
The device determines the difference between the last calibration and the penultimate
calibration, and issues a warning or an alarm depending on the setting configured. The
difference is an indicator for the condition of the sensor. Increasing differences indicate the
formation of buildup on the cathode. Clean the cathode as specified in the instructions in the
sensor operating manual.
Number of cap calibrations
Process check system
--> "Inputs: General/Frequently occurring functions" section ä 108
Path: Menu/Setup/Inputs/Chlorine/Extended setup/Diagnostics settings
Function Options Info
Delta zero point 0.0 to 3.2 nA Specify your limit values for monitoring the zero point
differential.
Function Options
•On
•Off
Factory setting
On
Warning limit Factory setting
1.0 nA
Diagnostics code and associated message text:
520 "Sensor calib."
Path: Menu/Setup/Inputs/Chlorine/Extended setup/Diagnostics settings
Function Options Info
Number of cap calibrations Specify how many calibrations may be performed with a
membrane cap before the cap has to be replaced.
The number depends heavily on the process and must be
determined individually.
Function Options
•On
•Off
Factory setting
Off
Warning limit 1 to 20
Factory setting
6
Diagnostics code and associated message text:
535 "Sensor check"