Liquistation CSF48 Inputs
27
Operation Options
•No operation
• Start sampling
•Program start
•Program stop
• Program duration
•Program pause
• Partprogram activation
•Change bottle
• Bottle synchronization
• External hold
• Start cleaning
Factory setting
No operation
No operation:
No action is executed.
Start sampling:
A pulse triggers sampling.
Program start:
A pulse starts a program.
Program stop:
A pulse stops the program running.
Program duration:
A program is active as long as the input signal is
present.
The signal is a level signal, i.e. the action takes effect as
long as the level is present. The level that triggers the
action is configured in the Signal slope menu item that
follows.
Program pause:
The input signal stops the program running. The
programs continue running when the signal
disappears.
The signal is a level signal, i.e. the action takes effect as
long as the level is present. The level that triggers the
action is configured in the Signal slope menu item that
follows.
Partprogram activation:
A pulse triggers a subprogram.
Change bottle:
A pulse triggers a changeover to the next bottle.
Bottle synchronization:
A pulse triggers a changeover to the set bottle position.
--> Then select the bottle position (depends on the
bottle configuration).
External hold:
The input signal triggers an external hold.
The signal is a level signal, i.e. the action takes effect as
long as the level is present. The level that triggers the
action is configured in the Signal slope menu item that
follows.
Start cleaning:
A pulse triggers the cleaning.
Signal slope Options
•Low-High
•High-Low
Factory setting
Low-High
Preselect the level change of the signal.
--> If Low-High is selected, the high level brings about
the corresponding setting.
If Input mode Armature end position detection is selected (only for version with sampling assembly):
Path: Menu/Setup/Inputs
Function Options Info