EasyManua.ls Logo

Endress+Hauser Liquistation CSF48 - Page 34

Endress+Hauser Liquistation CSF48
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programming Liquistation CSF48
34
The following table explains the various types of sampling using specific examples.
Type of sampling Example Info
Time-paced Sampling interval: 5 min
Sampling volume: 50 ml
•Bottle change mode: 2 h
With this setting, a 50 ml sample is taken every
5 minutes. 12 samples are thus taken every
hour. Each bottle is filled over a period of 2
hours. This results in a total sampling volume of
24 samples per bottles x 50 ml = 1200 ml.
This type of sampling remains constant over
time and does not take changes in flow or pollu-
ting load into account. It is possible to take a
representative sample if the intervals are brief
(e.g. 5 min).
Flow-paced Controlled via current input
Signal: 0 to 20 mA = 0 to 600 m³/h
Sampling volume: 50 ml
Sampling interval: 20 m³
•Bottle change mode: 2 h
If 20 mA = 600 m³/h, a sample is taken every 2
minutes (smallest sampling interval with maxi-
mum flow). The total number of samples
amounts to 60 samples per bottle. With a flow
of 300 m³/h, a sample is taken every 4 minutes.
Controlled via binary input
Signal pulse: 5 m³
Sampling volume: 50 ml
Sampling interval: 20 m³
•Bottle change mode: 2 h
The pulses are scaled at the flowmeter. By mul-
tiplying the pulses at the sampling interval, the
shortest sampling interval at the maximum
pulse frequency can be set.
Example: Given a maximum flow of 600 m³/h,
the pulse frequency at 5 m³ is 120 pulses/h or 2
pulses/min. With a sampling interval of 20 m³,
sampling takes place after 4 pulses = 2 minutes.
The current inputs can be configured for the
current range of 0 to 20 mA or 4 to 20 mA.
The binary inputs require power (24 V DC)
for floating contacts.
In the case of flow-paced sampling, the samp-
ling interval is calculated on the basis of the
volume flow. The same sampling volume is take
at variable intervals.
Advantage:
Good, representative results in the event of
minor fluctuations in flow.
Disadvantage:
Longer intervals when the level of water is low
mean that malfunctions cannot be detected.

Table of Contents

Other manuals for Endress+Hauser Liquistation CSF48

Related product manuals