Programming Liquistation CSF48
46
Dosing volume (for version
with vacuum pump or
sampling assembly)
Sampling volume (for version
with peristaltic pump)
Vacuum pump:
20 to 350 ml
Peristaltic pump:
10 to 10000 ml
Sampling assembly:
10 to 1000 ml
Factory setting
Vacuum pump:
200 ml
Peristaltic pump:
100 ml
Sampling assembly:
200 ml
Set the dosing volume or the sampling volume.
The volume is taken from the setup in the
version with a vacuum pump or sampling
assembly.
The dosing accuracy and the repeatability of a
sample volume < 20 ml can vary, depending on
the specific application.
Multiplier
(only for version with vacuum
pump)
1 to 10
Factory setting
1
You can change the sampling volume using the
multiplier.
For example, if the dosing volume is set to 200 ml, set
the multiplier to 2 for a sampling volume of 400 ml. 2
samples are taken in succession when sampling.
Bottle change mode Options
Number of samples
•Time
•External signal
Factory setting
Number of samples
The bottle can be changed after a specific number of
samples, after a time or by an external signal.
If Bottle change mode Number of samples is selected:
Samples per bottle 1 to 9999
Factory setting
1
Set the number of samples.
If Bottle change mode Time is selected:
Time interval 00-00:02 to 31-00:00
DD-HH:MM
Factory setting
00-01:00 DD-HH:MM
Set the time (in days, hours and minutes) after which
the system should change to the next bottle.
Multiple bottles 0 to 23
The configuration
options depend on the
current number of
bottles.
Factory setting
0
Multiple bottles:
"Simultaneous" transfer of two samples to separate
bottles.
Path: Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info