Programming Liquistation CSF48
48
4.2.3 Settings with a time/flow-paced Basic program (only for version with peristaltic
pump)
Settings with the Basic program type with 1 bottle
Sampling mode = "Time/flow paced CTVV"
Assignment bin. output Options
No binary output config. for
state reporting
Binary output S:x
Factory setting
No binary output config. for
state reporting
Assignment of the binary output to the program cycle.
Inputs Settings for the inputs can be made as
described in the "Inputs" section.
Path: Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Sampling volume input Options
No flow input configured
Binary input S:x
Current input S:x
Factory setting
No flow input configured
Select the sampling volume input.
The binary input or the current input must be
configured for this function.
Only the inputs configured as a sampling volume
input are displayed.
Sampling interval 00:01:00 to 99:59:00
HH:MM:SS
Factory setting
00:10:00 HH:MM:SS
Set the sampling interval.
If Sampling volume input Binary input is selected:
Sampling volume / pulse 10 to 1000 ml
Factory setting
20 ml
Set what sampling volume should be grabbed per
pulse.
The dosing accuracy and the repeatability of a
sample volume < 20 ml can vary, depending on
the specific application.
If Sampling volume input Current input is selected:
Sampling volume 20mA 10 to 10000 ml
Factory setting
100 ml
Set what sampling volume should be grabbed at 20
mA.
The dosing accuracy and the repeatability of a
sample volume < 20 ml can vary, depending on
the specific application.
Path: Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info