Liquistation CSF48 Programming
51
If Sampling volume input Current input is selected:
Sampling volume 20mA 10 to 10000 ml
Factory setting
100 ml
Set what sampling volume should be grabbed at 20
mA.
The dosing accuracy and the repeatability of a
sample volume < 20 ml can vary, depending on
the specific application.
Flow calculation Options
Current
•Average flow
Factory setting
Current
Current:
The current flow is converted to the sampling volume
at the time of sampling.
Average flow:
The system calculates the mean between the last and
the current sample and sets the sampling volume
accordingly.
Bottle change mode Options
•Number of samples
•Time
•External signal
Factory setting
Number of samples
The bottle can be changed either after a specific
number of samples, after a time or by an external
signal.
If Bottle change mode Number of samples is selected:
Samples per bottle 1 to 9999
Factory setting
1
Set the number of samples.
If Bottle change mode Time is selected:
Time interval 00-00:02 to 31-00:00
DD-HH:MM
Factory setting
00-01:00 DD-HH:MM
Set the time (in days, hours and minutes) after which
the system should change to the next bottle.
Multiple bottles 0 to 23
The configuration
options depend on the
current number of
bottles
Factory setting
0
Multiple bottles:
"Simultaneous" transfer of two samples to separate
bottles.
If Bottle change mode External signal is selected:
Bottle change signal input Options
No bottle change input
configured
Binary input S:x
Factory setting
No bottle change input
configured
The bottle change input can be configured under
Inputs .
Path: Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info