Programming Liquistation CSF48
58
Start volume input Options
• No flow input configured
• Binary input S:x
• Current input S:x
Factory setting
No flow input configured
Select the start volume input.
The binary input or the current input must be confi-
gured for this function.
Only the inputs configured for flow measurement are
displayed.
Start flow sum 1.000 to 9999.000 m³
Factory setting
10.000 m³
Set the start volume.
Stop condition Options
•Program end
• Continuous
•Date/time
Factory setting
Program end
Program end:
The device stops sampling automatically once it has
run through the set program. All the assigned bottles
are filled.
Continuous:
The device runs through the set program conti-
nuously in an infinite loop. Do not forget to empty the
bottles.
Date/time:
The device stops the set program at a specific time.
If Stop condition Date/time is selected:
Stop date 01.01.2000 to 31.12.2099
Factory setting
DD.MM.YYYY
Set the stop date for the sampling program. The for-
mat depends on the option configured under general
settings.
Stop time 00:00:00 to 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is stopped.
The format depends on the option configured under
general settings.
Setup subprogram
New
Programpart Use a distinct name for your subprogram. The pro-
gram name can be up to 16 characters long.
Sampling mode Options
•Time paced CTCV
•Flow paced VTCV
• Time/flow paced CTVV
•External signal
Factory setting
Time paced CTCV
Time paced CTCV:
A constant sampling volume is taken at steady inter-
vals.
Flow paced VTCV:
A constant sampling volume is taken at variable
intervals.
Time/flow paced CTVV(only for version with peris-
taltic pump):
A variable sampling volume is taken at steady inter-
vals.
External signal
A pulse at the binary input starts a sampling cycle
Path: Menu/Setup/Sampling programs
Function Options Info