Liquistation CSF48 Mounting
Endress+Hauser 21
For version with sampling assembly
A0024412
11 Installation conditions for Liquistation CSF48 with Samplefit CSA420 sampling assembly
Note the following when installing the sampling assembly in a pipe:
• The best installation location is in the ascending pipe (pos. 2). Installation is also possible in
the horizontal pipe (pos. 1).
• Avoid installation in the down pipe (pos. 4).
• Avoid siphoning effects in the sample line.
• The minimum vertical distance between the assembly and the inlet of the sampler should
be at least 0.5 m (1.65 ft).
Note the following when erecting the sampler:
• Erect the device on a level surface.
• Protect the device against additional heating (e.g. from a heating system).
• Protect the device against mechanical vibrations.