Liquistation CSF48 Mounting
Endress+Hauser 21
For version with sampling assembly
A0024412
 11 Installation conditions for Liquistation CSF48 with Samplefit CSA420 sampling assembly
Note the following when installing the sampling assembly in a pipe:
The best installation location is in the ascending pipe (pos. 2). Installation is also possible in
the horizontal pipe (pos. 1).
Avoid installation in the down pipe (pos. 4).
Avoid siphoning effects in the sample line.
The minimum vertical distance between the assembly and the inlet of the sampler should
be at least 0.5 m (1.65 ft).
Note the following when erecting the sampler:
Erect the device on a level surface.
Protect the device against additional heating (e.g. from a heating system).
Protect the device against mechanical vibrations.